TENS & EMS Combo, 5 modes of TENS, 3 Modes of EMS, 100mA output, Hard Carrying Case
|
Your Price: Only available to healthcare practitioners, healthcare verified log in account required to purchase
Item Number: TENS/EMS ePower® TE2 Combo
Manufacturer: ePower®
Manufacturer Part No: TENS/EMS ePower® TE2 Combo
|
The ePower® TE2 TENS/EMS Combo gives the user the choice of 5 modes of TENS and 2 Modes of EMS with full strength 100mA output.
System Includes: Stimulator, Lead Wires, 1 package of Self-Adhesive Electrodes, 9V Battery, Instruction Booklet, Hard Carrying Case.
Specifications:
Power sources: One 9 volt battery
Output channel: Dual channel
Waveform: Bi-phase square-wave pulse
Output current: 0-50V (Load:500 ohm)
Output intensity: adjustable
Treatment type: TENS and EMS mode
P.W. (Pulse Width): Adjustable, from 50 to 300 μs, 10 μs/step
P.R. (Pulse Rate): Adjustable, from 2 to 150 Hz, 1 Hz/step
Timer: Adjustable, from 1 to 60 minutes or continuous.
On Time: Adjustable, 2~90 seconds,1Sec./step.
Off Time: Adjustable, 0~90 seconds,1Sec./step.
Ramp Time: Adjustable, 1-8 seconds, 1 Sec./step.
Modes: B (Burst), N (Normal), M (Modulation) SD1
(Strength Duration) SD2 ,S(Synchronous) and
A(Alternate).
Burst Mode: Burst rate: adjustable, 0.5 - 5 Hz,
Pulse width: adjustable, 50~300 μs
Frequency: fixed = 100 Hz
Normal Mode: The pulse rate and pulse width are adjustable. It
generates continuous stimulation based on the
setting value.
Modulation Mode: Modulation mode is a combination of pulse
rate and pulse width modulation. The pulse
rate and width are automatically varied in a
cycle pattern. The pulse width is decreased by
50% from its original setting in the 0.5 second,
then the pulse rate is decreased by 50% from
its original setting in the 0.5 second. Total
cycle time is 1 second. In this mode, pulse rate
(2~150Hz) and pulse width (50-300μs) are fully
adjustable.
SD1 Mode: The SD1(Strength-Duration) mode consists
of automatic modulation intensity and pulse
width in 40% range. The intensity is always
increasing while the pulse width is decreasing
and vice versa. The intensity is decreased by
40% while the pulse width is increased by 40%
in 5 seconds. In the next 5 seconds, the intensity
is increased by 40% while the pulse width
is decreased by 40%. Total cycle time is 10
seconds. Pulse rate (2~150Hz) and pulse width
(50-300μs) are fully adjustable.
15
SD2 Mode: The SD2 (Strength-Duration) mode consists
of automatic modulation intensity and pulse
width in 70% range. The intensity and pulse
width in 70% range. The intensity is always
increasing while the pulse width is decreasing
and vice versa. The intensity is decreased by
70% while the pulse width is increasing by 70%
in 5 seconds. In the next 5 seconds, the intensity
is increased by 70% while the pulse width
is decreased by 70%. Total cycle time is 10
seconds. Pulse rate (2~150Hz) and pulse width
(50-300μs) are fully adjustable.
Synchronous
mode:
S t i m u l a t i o n o f b o t h c h a n n e l s o c c u r s
synchronously. The “ON” time including
R a m p U p ” a n d “ R a m p D o w n ” t i m e .
Therefore, the setting of ON Time should be no
less than two times of the “Ramp” time in
this mode.
ON TIME ≥ Ramp up + Ramp down
Alternate Mode: The stimulation of the CH2 will occur after the
1st contraction of CH1 is completed. In this
mode, the setting of ON Time should be no less
than two times of the “Ramp” time.
The OFF Time should be equal or more than the
ON Time.
ON TIME ≥ Ramp up + Ramp down
OFF TIME ≥ ON TIME
Patient Compliance Meter: This unit can store 60 sets of operation records. Total recorded time is 999 hours.
Low Battery Indication: A low battery indicator will show up on the LCD when the battery is low
|
|
|
|